Jalalian E, Hashemi E, Naser Mostufi S, Banifatemeh A, Shariati M, Pirmoazen A et al . Effect of Abutment Connection Type and Cyclic Loading on Removal Torque Value. J Res Dent Maxillofac Sci 2019; 4 (1) :36-40
URL: http://jrdms.dentaliau.ac.ir/article-1-243-en.html
1- Associate Professor, Prosthodontics Dept, Faculty of Dentistry, Tehran Medical Sciences,
2- Postgraduate Student, Prosthodontics Dept, Faculty of Dentistry, Tehran Medical Sciences, Islamic Azad University, Tehran, Iran , Hashemi.prostho@gmail.com
3- Assistant professor, Prosthodontics Dept, Faculty of Dentistry, Tehran Medical Sciences,
4- Postgraduate Student, Prosthodontics Dept, Faculty of Dentistry, Tehran Medical Sciences,
5- Prosthodontics
Abstract:   (3212 Views)

Background and Aim: Maintaining the primary torque of the abutment screw is a common problem related to implant-supported restorations; a failure in this respect often leads to screw loosening. The abutment screw plays a critical role in the long-term stability of restorations. This study aimed to assess the effect of two abutment connection types with and without cyclic loading on removal torque value (RTV).
Materials and Methods: A total of 24 abutments and fixtures were evaluated in this in-vitro experimental study. Two abutments with internal and external connections were tightened to fixtures using a digital torque-meter according to the manufacturer's instructions. Half of the samples were subjected to cyclic loading with a 100-N mechanical load at a 1-Hz frequency (500,000 cycles) applied at 23±1°C with 50±5% humidity. Next, the RTV was measured for each abutment. Two-way analysis of variance (ANOVA) was used to assess the effect of the abutment system and cyclic loading on RTV. 
Results: The highest RTV belonged to the internal connection abutment without cyclic loading (22 Newton-centimeter (Ncm)), while the lowest RTV was noted with the external connection abutment subjected to cyclic loading (14 Ncm; P<0.01). There was no statistically significant interaction effect between the two variables (the abutment system and cyclic loading; P=0.839). The RTV in the internal hexagon abutments was significantly higher than that in the external hexagon abutment system (P<0.001). Cyclic loading significantly reduced the RTVs in both abutments (P<0.001).  
Conclusion: According to the results, internal connection abutments experience less screw loosening.
 

Full-Text [PDF 338 kb]   (1632 Downloads) |   |   Full-Text (HTML)  (1336 Views)  
Type of Study: Original article | Subject: Dental implant

References
1. Chaar MS, Att W, Strub JR. Prosthetic outcome of cement‐retained implant‐supported fixed dental restorations: a systematic review. J Oral Rehabil. 2011 Sep;38(9):697-711. [DOI:10.1111/j.1365-2842.2011.02209.x] [PMID]
2. Nassar HI, Abdalla MF. Bacterial leakage of different internal implant/abutment connection. Future Dent J. 2015 Dec 1;1(1):1-5. [DOI:10.1016/j.fdj.2015.09.001]
3. Ricci G, Aimetti M, Stablum W, Guasti A. Crestal bone resorption 5 years after implant loading: clinical and radiologic results with a 2-stage implant system. Int J Oral Maxillofac Implants. 2004 Jul-Aug;19(4):597-602.
4. Goodacre CJ, Bernal G, Rungcharassaeng K, Kan JY. Clinical complications with implants and implant prostheses. J Prosthet Dent. 2003 Aug 1;90(2):121-32. [DOI:10.1016/S0022-3913(03)00212-9]
5. Johansson LÅ, Ekfeldt A. Implant-supported fixed partial prostheses: a retrospective study. Int J Prosthodont. 2003 Mar-Apr;16(2):172-6.
6. Byrne D, Jacobs S, O'Connell B, Houston F, Claffey N. Preloads generated with repeated tightening in three types of screws used in dental implant assemblies. J Prosthodont. 2006 May-Jun;15(3):164-71. [DOI:10.1111/j.1532-849X.2006.00096.x] [PMID]
7. Hebel KS, Gajjar RC. Cement-retained versus screw-retained implant restorations: achieving optimal occlusion and esthetics in implant dentistry. J Prosthet Dent. 1997 Jan 1;77(1):28-35. [DOI:10.1016/S0022-3913(97)70203-8]
8. Ribeiro CG, Maia ML, Scherrer SS, Cardoso AC, Wiskott HW. Resistance of three implant-abutment interfaces to fatigue testing. J Appl Oral Sci. 2011 Aug;19(4):413-20. [DOI:10.1590/S1678-77572011005000018] [PMID] [PMCID]
9. Ricomini Filho AP, Fernandes FS, Straioto FG, da Silva WJ, Del Bel Cury AA. Preload loss and bacterial penetration on different implant-abutment connection systems. Braz Dent J. 2010;21(2):123-9. [DOI:10.1590/S0103-64402010000200006] [PMID]
10. Harder S, Dimaczek B, Açil Y, Terheyden H, Freitag-Wolf S, Kern M. Molecular leakage at implant-abutment connection-in vitro investigation of tightness of internal conical implant-abutment connections against endotoxin penetration. Clin Oral Investig. 2010 Aug 1;14(4):427-32. [DOI:10.1007/s00784-009-0317-x] [PMID]
11. Lang LA, Wang RF, May KB. The influence of abutment screw tightening on screw joint configuration. J Prosthet Dent. 2002 Jan 1;87(1):74-9. [DOI:10.1067/mpr.2002.121488] [PMID]
12. Vigolo P, Majzoub Z, Cordioli G. Measurement of the dimensions and abutment rotational freedom of gold-machined 3i UCLA-type abutments in the as-received condition, after casting with a noble metal alloy and porcelain firing. J Prosthet Dent. 2000 Nov ;84(5):548-53. [DOI:10.1067/mpr.2000.110497] [PMID]
13. Theoharidou A, Petridis HP, Tzannas K, Garefis P. Abutment screw loosening in single-implant restorations: a systematic review. Int J Oral Maxillofac Implants. 2008 Jul-Aug;23(4):681-90.
14. Malaguti G, Denti L, Bassoli E, Franchi I, Bortolini S, Gatto A. Dimensional tolerances and assembly accuracy of dental implants and machined versus cast‐on abutments. Clin Implant Dent Relat Res. 2011 Jun;13(2):134-40. [DOI:10.1111/j.1708-8208.2009.00189.x] [PMID]
15. Cashman PM, Schneider RL, Schneider GB, Stanford CM, Clancy JM, Qian F. In vitro analysis of post-fatigue reverse-torque values at the dental abutment/implant interface for a unitarian abutment design. J Prosthodont. 2011 Oct;20(7):503-9. [DOI:10.1111/j.1532-849X.2011.00756.x] [PMID]
16. Siadat H, Pirmoazen S, Beyabanaki E, Alikhasi M. Does abutment collar length affect abutment screw loosening after cyclic loading? J Oral Implantol. 2015 Jul;41(S1):346-51. [DOI:10.1563/AAID-JOI-D-D-14-00021] [PMID]
17. Kim ES, Shin SY. Influence of the implant abutment types and the dynamic loading on initial screw loosening. J Adv Prosthodont. 2013 Feb 1;5(1):21-8. [DOI:10.4047/jap.2013.5.1.21] [PMID] [PMCID]
18. Yeo IS, Lee JH, Kang TJ, Kim SK, Heo SJ, Koak JY, et al. The effect of abutment screw length on screw loosening in dental implants with external abutment connections after thermocycling. Int J Oral Maxillofac Implants. 2014 Jan-Feb;29(1):59-62. [DOI:10.11607/jomi.3125] [PMID]
19. Pintinha M, Camarini ET, Sábio S, Pereira JR. Effect of mechanical loading on the removal torque of different types of tapered connection abutments for dental implants. J Prosthet Dent. 2013 Nov;110(5):383-8. [DOI:10.1016/j.prosdent.2013.06.007] [PMID]
20. Sailer I, Sailer T, Stawarczyk B, Jung RE, Hämmerle CH. In vitro study of the influence of the type of connection on the fracture load of zirconia abutments with internal and external implant-abutment connections. Int J Oral Maxillofac Implants. 2009 Sep-Oct;24(5):850-8.
21. Maeda Y, Satoh T, Sogo M. In vitro differences of stress concentrations for internal and external hex implant-abutment connections: a short communication. J Oral Rehabil. 2006 Jan;33(1):75-8. [DOI:10.1111/j.1365-2842.2006.01545.x] [PMID]
22. Piermatti J, Yousef H, Luke A, Mahevich R, Weiner S. An in vitro analysis of implant screw torque loss with external hex and internal connection implant systems. Implant Dent. 2006 Dec 1;15(4):427-35. [DOI:10.1097/01.id.0000245440.09464.48] [PMID]
23. Tsuge T, Hagiwara Y. Influence of lateral-oblique cyclic loading on abutment screw loosening of internal and external hexagon implants. Dent Mater J. 2009;28(4):373-81. [DOI:10.4012/dmj.28.373] [PMID]
24. Junqueira MC, Ribeiro RF, Faria AC, Macedo AP, Almeida RP. Screw loosening of different UCLA-type abutments after mechanical cycling. Braz J Oral Sci. 2013 Sep;12(3):228-32. [DOI:10.1590/S1677-32252013000300014]
25. Kitagawa T, Tanimoto Y, Odaki M, Nemoto K, Aida M. Influence of implant/abutment joint designs on abutment screw loosening in a dental implant system. J Biomed Mater Res B Appl Biomater. 2005 Nov;75(2):457-63. [DOI:10.1002/jbm.b.30328] [PMID]
26. Gracis S, Michalakis K, Vigolo P, Vult von Steyern P, Zwahlen M, Sailer I. Internal vs. external connections for abutments/reconstructions: a systematic review. Clin Oral Implants Res. 2012 Oct;23:202-16. [DOI:10.1111/j.1600-0501.2012.02556.x] [PMID]
27. Kwon JH, Han CH, Kim SJ, Chang JS. The change of rotational freedom following different insertion torques in three implant systems with implant driver. J Adv Prosthodont. 2009 Mar 1;1(1):37-40. [DOI:10.4047/jap.2009.1.1.37] [PMID] [PMCID]
28. Cho WR, Huh YH, Park CJ, Cho LR. Effect of cyclic loading and retightening on reverse torque value in external and internal implants. J Adv Prosthodont. 2015 Aug;7(4):288-93. [DOI:10.4047/jap.2015.7.4.288] [PMID] [PMCID]
29. Feitosa PC, de Lima AP, Silva-Concílio LR, Brandt WC, Neves AC. Stability of external and internal implant connections after a fatigue test. Eur J Dent. 2013 Jul;7(3):267. [DOI:10.4103/1305-7456.115407] [PMID] [PMCID]
30. Tsuruta K, Ayukawa Y, Matsuzaki T, Kihara M, Koyano K. The influence of implant-abutment connection on the screw loosening and microleakage. Int J Implant Dent. 2018 Apr 9;4(1):11. [DOI:10.1186/s40729-018-0121-y] [PMID] [PMCID]
31. Zarbakhsh A, Mazaheri Tehrani A, Shamshirgar F, Khosroshahi H. Effect of GapSeal® as a Sealing Material on Microgap and Microleakage at External Hexagon Implant Connections Following Cyclic Loading: An In Vitro Study. J Res Dentomaxillofac Sci. 2018;3(3):42-48.
32. Benjaboonyazit K, Chaijareenont P, Khongkhunthian P. Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study. Int J Implant Dent. 2019 Jan 14;5(1):1. [DOI:10.1186/s40729-018-0154-2] [PMID] [PMCID]

Add your comments about this article : Your username or Email:
CAPTCHA

Send email to the article author


Rights and permissions
Creative Commons License This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.

© 2024 CC BY-NC 4.0 | Journal of Research in Dental and Maxillofacial Sciences

Designed & Developed by: Yektaweb