Shahbazi S, Moezzi ghadim N, Mirzaei A, Azizi A. Comparison of Soft Tissue Thermal Changes Induced by Three Types of Diode Lasers at 810, 940, and 980nm Wavelengths. J Res Dent Maxillofac Sci 2020; 5 (2) :7-13
URL: http://jrdms.dentaliau.ac.ir/article-1-269-en.html
1- Dentist
2- Assistant professor, Oral Medicine Dept, Faculty of Dentistry,Tehran Medical Sciences
3- Educational Trainer at International Academy of RWTH Aachen University, Aachen, Germany
4- Professor,Oral Medicine Dept, Faculty of Dentistry,Tehran Medical Sciences , Islamic Azad University , Tehran, Iran. , drarashazizi@yahoo.com
Full-Text [PDF 323 kb]   (1105 Downloads)     |   Abstract (HTML)  (2566 Views)
Full-Text:   (938 Views)


Abstract


Background and Aim: Diode laser is a great choice for soft tissue surgery. The diode laser is partially absorbed by hard dental tissues, making it safe for soft tissue surgery. This study aimed to compare the tissue thermal changes induced by three types of diode lasers at 810, 940, and 980nm wavelengths.
Materials and Methods: In this in-vitro experimental study, using a diode laser device (continuous mode with a 400μm fiber tip) at each of the 810, 940, and 980nm wavelengths and powers of 1, 2, and 3W in contact mode, incisions with the length of 2cm were made on pieces of meat over a period of 10 seconds. The primary and secondary temperatures were measured using a thermocouple. Data were analyzed using two-way analysis of variance (ANOVA).
Results: Tissue temperature changes induced by diode laser at 810nm/3W were significantly greater than that of 2W power. These changes were higher with 2W of power compared to 1W (P<0.05). Temperature changes induced by diode laser at 940nm/2 and 3W were significantly greater than that of 1W power (P<0.05). Temperature changes induced by diode laser at 980nm/3W were significantly greater than that of 2W power. Tissue temperature changes were higher with 2W of power compared to 1W (P<0.05).
Conclusion: Diode laser (continuous mode with a 400µm fiber tip) at 3W of power and 980nm wavelength caused the highest rate of thermal changes. The 810nm diode laser with the power of 1W caused the slightest heat changes in the soft tissue.
Keywords: Diode Laser, Temperature, Soft Tissue Injuries
 

Introduction


The surgical blade has been traditionally used in soft tissue surgery for many years. The surgical blade is the first choice in soft tissue surgery due to its accuracy, speed, ease of use, reasonable cost, and minimal damage to the tissues surrounding the surgical site. However, it cannot provide a bleeding-free environment necessary for surgeries of highly vascularized tissues. (1,2)
In contrast, lasers are a new method introduced for oral soft tissue surgery. The main difference between surgical blades and laser is the ability to create homeostasis by the laser. (2,3) Other benefits of laser include high cutting power, disinfection of the surgical environment, better vision, minimal bleeding, reduced edema, infection, postoperative pain, and scarring, as well as faster healing and increased patient acceptance. (1-4) But this method also has limitations, such as the high cost of equipment, the training of specialist personnel for the clinical applications of different lasers, the need for more than one laser (with different wavelengths) in diverse processes, and thermal damage. (5) Lasers used in soft tissue surgery include erbium-doped yttrium aluminum garnet (Er:YAG), carbon dioxide (CO2), neodymium-doped yttrium aluminum garnet (Nd:YAG), diode, and erbium, chromium-doped yttrium-scandium-gallium-garnet (Er,Cr:YSGG) lasers. (6,7) The diode laser is a great choice for soft tissue surgery. The diode laser is partially absorbed by hard dental tissues, thereby making it safe for soft tissue surgery; it is now widely used in surgical excision of oral pathologic lesions. (5) The biggest advantage of a diode laser is that it is a portable device of small size and low weight. (5,8) The diode laser causes horizontal and vertical thermal damage around the incision site, which results in carbonization, necrosis, and irreversible tissue changes, delaying the wound healing, reducing the tensile strength of the tissue, causing pulp and periodontium tissue injury, causing delay in graft repair, and preventing the correct histopathological diagnosis of the specimens. (1,4,6)
Various studies have compared the tissue thermal changes caused by different lasers, but none of the previous studies has investigated the effect of different wavelengths of diode laser (810, 940, and 980nm) on soft tissue thermal changes. Considering the information gap mentioned above, in this study, it is intended to compare the soft tissue thermal changes induced by three types of diode lasers at 810, 940, and 980nm wavelengths in the Faculty of Dentistry of Islamic Azad University of Medical Sciences, Tehran, Iran.


Materials and Methods


In this in-vitro experimental study, 2 kg of lean beef was prepared 24 hours before the study. Then, the meat was divided into 4×4×2 cm3 pieces and was kept in the refrigerator at 4°C. One hour before laser irradiation, the meat was removed from the refrigerator to make the temperature similar to the environment. To uniformize the temperature of the meat, the primary temperature of the meat was measured using a thermocouple (D55 Hanyoung Nux, South Korea). The purchased thermocouples were placed one millimeter (mm) from the laser irradiation site at a depth of five mm and perpendicularly at the center of the incisions (Figure 1). (7) To uniformize the incisions, templates prepared with the dimensions of 2×2 cm2 were used. Then, using a diode laser device (continuous mode with a 400μm fiber tip) at each of the 810, 940, and 980nm wavelengths and powers of 1, 2, and 3W in contact mode, incisions with the length of 2cm were made on the pieces of meat over a period of 10 seconds (cutting speed of 2mm/second; Figure 2) as follows:
The wavelength of 810nm (Pulsar, Iran): the power of 1W (4 times) - 2W (4 times) - 3W (4 times)
The wavelength of 940nm (Biolase Technology, Irvine, CA, USA): the power of 1W (4 times) - 2W (4 times) - 3W (4 times)
The wavelength of 980nm (Wiser Doctor Smile, Lambda SPA, Italy): the power of 1W (4 times) - 2W (4 times) - 3W (4 times).
The secondary temperature shown by the thermocouple after laser irradiation was recorded in datasheets.
Data were analyzed using two-way analysis of variance (ANOVA).

Figure 1. Meat preparation and thermocouple placement to measure the initial temperature
Figure 2. 810, 940, and 980nm lasers used in research



Results


In this in-vitro experimental study, using a diode laser device (continuous mode with a 400μm fiber tip) at each of the 810, 940, and 980nm wavelengths and the powers of 1,2, and 3W in contact mode, 2cm incisions (1mm/second cutting speed) were made on 36 pieces of meat. 
The primary and secondary temperatures indicated by the thermocouples before and after laser irradiation were recorded in datasheets, and the resulting thermal changes were examined. The results are summarized in Tables 1 to 9 and Figure 3.
With the 810nm wavelength and power of 1W, the average tissue temperature increased from 19.7 to 20.45 (0.75°C). This temperature increased from 19.7 to 21.3 (1.6°C) with 2W of power and from 19.7 to 22.45 (2.75°C) with 3W of power. Tissue temperature changes induced by diode laser at 810nm/3W were significantly greater than that of 2W power. Tissue temperature changes were higher with 2W of power compared to 1W (P<0.05; Tables 1 and 4).       
Table 1: Tissue temperature changes induced by diode laser at 810nm wavelength and different powers

With the 940nm wavelength and power of 1W, the average tissue temperature increased from 18.4 to 19.2 (0.8°C). This temperature increased from 19.8 to 22.32 (2.52°C) with 2W of power and from 20 to 22.8 (2.8°C) with 3W of power. Tissue temperature changes induced by diode laser at 940nm/2 and 3W were significantly greater than that of 1W power (P<0.05; Tables 2 and 5).

Table 2: Tissue temperature changes induced by diode laser at 940nm wavelength and different powers

With the 980nm wavelength and power of 1W, the average tissue temperature increased from 20.2 to 21.1 (0.9°C). This temperature increased from 18.4 to 20.17 (2.77°C) with 2W of power and from 19.9 to 24.57 (4.67°C) with 3W of power. Tissue temperature changes induced by diode laser at 980nm/3W were significantly greater than that of 2W power. Tissue temperature changes were greater with 2W of power compared to 1W (P<0.05; Tables 3 and 6).

Table 3: Tissue temperature changes induced by diode laser at 980nm wavelength and different powers

Tissue temperature changes induced by diode laser at 810nm/3W were significantly greater than that of 2W power. Tissue temperature changes were higher with 2W of power compared t
o 1W (Table 4).
Table 4: Tissue temperature changes induced by diode laser at 810nm wavelength and different powers

Tissue temperature changes induced by diode laser at 940nm/2 and 3W were significantly greater than that of 1
W power (Table 5).
Table 5: Tissue temperature changes induced by diode laser at 940nm wavelength and different powers

Tissue temperature changes induced by diode laser at 980nm/3W were significantly greater than that of 2W power. Tissue temperature changes were greater with 2W of power compared to
1W (Table 6).
Table 6: Tissue temperature changes induced by diode laser at 980nm wavelength and different powers

The tissue thermal changes caused by the diode laser with 1W power at 810, 940, and 980nm wavelengths were not significantly different (Table 7).


Table 7: Tissue temperature changes induced by the diode laser with 1W power at different wavelengths

The tissue thermal changes induced by the diode laser with 2W power
at 980nm were significantly greater than those at the 810nm wavelength were. The 940nm wavelength showed no significant difference with 980nm and 810nm wavelengths (Table 8).

Table 8: Tissue temperature changes induced by the diode laser with 2W power at different wavelengths
The tissue thermal changes induced by the diode laser with 3W
power at 980 nm were significantly greater than those at 810nm and 940nm wavelengths (Table 9).

Table 9: Thermal changes induced by the diode laser with 3W power at different wavelengths

Figure 3. Tissue thermal changes induced by diode laser at 810, 940, and 980nm wavelengths with 1, 2, and 3W of power; WL=Wavelength, CI=Confidence Interval

 

Discussion


According to the experiments and the results obtained, the soft tissue thermal changes induced by the diode laser at 810nm wavelength with 3W power were significantly greater than that with the power of 2W. These changes were significantly greater with the power of 2W compared to 1W. With the 940nm wavelength with the powers of 2W and 3W, the changes were significantly greater than that at 1W, and with the 980nm wavelength with the power of 3W, the changes were significantly greater than that at 2W. These changes at 2W were significantly greater than those at 1W were.
In addition, tissue thermal changes induced by the diode laser with 1W power at different wavelengths of 810, 940, and 980nm were not significantly different. With the power of 2W at a 980nm wavelength, these changes were significantly greater than that at the 810nm wavelength. Thermal changes were significantly greater at a 980nm wavelength with the power of 3W than that at 810nm and 940nm wavelengths
In 2012, Beer et al histologically examined the effects of different modes (pulse and micropulse) and powers (2.5, 3.5, and 4.5W) of a 980nm diode laser on soft tissue thermal damage of bovine liver samples. (1) They concluded that the depth and the width of the incision and the extent and the depth of carbonization, necrosis, and irreversible tissue damage were significantly and inversely correlated with the cutting speed (P<0.0001) and directly correlated with the power. The extent and the depth of the reversible injury area were inversely correlated with the power (P<0.01). The micropulse mode created a smaller area of ​​carbonization and necrosis and a smaller incision width, (1) similar to the results of the present study, in that with the diode laser at a wavelength of 980nm, with increasing the power, tissue thermal changes significantly increased.
In 2010, Cercadillo-Ibarguren et al investigated the thermal and histological effects of diode, Er,Cr:YSGG (830nm with 2, 5, and 10W, pulse mode), and CO2 lasers on the porcine soft tissue of oral mucosa. (3) They showed that the lowest thermal effects resulted from the Er,Cr:YSGG laser with 1W of power and air-water spray and then with the CO2 laser with a power of 20W in pulse mode and then with the diode laser with a power of 2W in pulse mode. (3) The results of the cited study are also consistent with our study that, with increasing the power, the temperature of the tissue significantly increases with the 810nm diode laser.
In 2004, Fontana et al studied thermal damage caused by a diode laser at a wavelength of 810nm and powers of 600mW, 800mW, 1W, and 1.2W in continuous mode in the mouse oral mucosa (in vitro) and the mandibular alveolar bone (in vivo) using an electronic thermistor. (8) In this case, the temperature raised to 2.8°C; this heat change is considered safe and close to thermal injury. (8) In the mentioned study, with increasing the power and duration of diode laser irradiation, thermal changes in different tissue thicknesses (1, 2, and 3mm) and periodontal pocket increased, which is in line with the findings of the current study.
In a study by Goharkhay et al in 1999, the cutting characteristics and damage of porcine mandibular soft tissue because of the application of an 810nm diode laser (pulse and continuous with 0.5-4.5W of power and 200 and 400µm fiber tips) were histologically evaluated. (9) The results of the said study showed that the depth and the width of the incisions directly and strongly correlated with the power, and no damage was observed in the bone beneath the incisions. (9) In the cited study, the extent of tissue damage correlated with laser parameters (pulse/continuous, time, power, fiber tip, etc.). (9,10)
In all studies on diode lasers, the thermal effects are directly proportional to the increase in power. At high powers, compared to low powers, heat accumulates mainly in the tissue until it spreads to the surrounding structures, which increases the temperature. (4) The reason for the increase in thermal changes with increasing the wavelength is that there is more water and hemoglobin absorption with the 980nm diode laser compared to the 810 and 940nm diode lasers. At higher wavelengths, due to the good absorption of these wavelengths by water and hemoglobin of tissues, a further increase in temperature is observed. (11)


Conclusion


The results of the present study showed that diode laser (continuous mode with a 400µm fiber tip) with 3W of power at a 980nm wavelength caused the highest rate of thermal changes in the soft tissue. Diode laser with the power of 1W at the wavelength of 810nm caused the slightest heat changes in the soft tissue.

Type of Study: Original article | Subject: Oral medicine

References
1. Beer F, Körpert W, Passow H, Steidler A, Meinl A, Buchmair AG, et al. Reduction of collateral thermal impact of diode laser irradiation on soft tissue due to modified application parameters. Lasers Med Sci. 2012 Sep;27(5):917-21. [DOI:10.1007/s10103-011-1007-x] [PMID]
2. Stübinger S, Saldamli B, Jürgens P, Ghazal G, Zeilhofer HF. Soft tissue surgery with the diode laser--theoretical and clinical aspects. Schweiz Monatsschr Zahnmed. 2006;116(8):812-20.
3. Cercadillo-Ibarguren I, España-Tost A, Arnabat-Domínguez J, Valmaseda-Castellón E, Berini-Aytés L, Gay-Escoda C. Histologic evaluation of thermal damage produced on soft tissue by CO2, Er,Cr:YGSS and diode lasers. Med Oral Patol Oral Cir Bucal. 2010 Nov 1;15(6):e912-8. [DOI:10.4317/medoral.15.e912] [PMID]
4. Wilder-Smith P, Arrastia AM, Liaw LH, Berns M. Incision properties and thermal effects of three CO2 lasers in soft tissue. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 1995 Jun;79(6):685-91. [DOI:10.1016/S1079-2104(05)80300-2]
5. David CM, Gupta P. Lasers in Dentistry: A Review. Int J Adv Health Sci. 2015 Dec;2(8):7-13.
6. Angiero F, Parma L, Crippa R, Benedicenti S. Diode laser (808 nm) applied to oral soft tissue lesions: a retrospective study to assess histopathological diagnosis and evaluate physical damage. Lasers Med Sci. 2012 Mar;27(2):383-8. [DOI:10.1007/s10103-011-0900-7] [PMID]
7. Parker S. Lasers and soft tissue: 'fixed' soft tissue surgery. Br Dent J. 2007 Mar 10;202(5):247-53. [DOI:10.1038/bdj.2007.172] [PMID]
8. Fontana CR, Kurachi C, Mendonça CR, Bagnato VS. Temperature variation at soft periodontal and rat bone tissues during a medium-power diode laser exposure. Photomed Laser Surg. 2004 Dec;22(6):519-22. [DOI:10.1089/pho.2004.22.519] [PMID]
9. Goharkhay K, Moritz A, Wilder-Smith P, Schoop U, Kluger W, Jakolitsch S, et al. Effects on oral soft tissue produced by a diode laser in vitro. Lasers Surg Med. 1999;25(5):401-6. https://doi.org/10.1002/(SICI)1096-9101(1999)25:5<401::AID-LSM6>3.0.CO;2-U [DOI:10.1002/(SICI)1096-9101(1999)25:53.0.CO;2-U]
10. Khalifeh S, Rezaee E, Azizi A, Ebrahimi H. Efficacy of Low-Level Laser Versus Transcutaneous Electrical Nerve Stimulation for Treatment of Myofascial Pain Syndrome: A Clinical Trial. J Res Dentomaxillofac Sci. 2018;3(3):33-41. [DOI:10.29252/jrdms.3.3.33]
11. Yao J, Liu B, Qin F. Rapid temperature jump by infrared diode laser irradiation for patch-clamp studies. Biophys J. 2009 May 6;96(9):3611-9. [DOI:10.1016/j.bpj.2009.02.016] [PMID] [PMCID]

Add your comments about this article : Your username or Email:
CAPTCHA

Send email to the article author


Rights and permissions
Creative Commons License This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.

© 2024 CC BY-NC 4.0 | Journal of Research in Dental and Maxillofacial Sciences

Designed & Developed by: Yektaweb